VitalCare Series Patient Monitors
MULTIPLE CONFIGURATIONS/ BATTERY MANAGEMENT/ COST EFFECTIVE
Recomended For Dental Offices, Oral Surgery, Medical and Surgical
Proprietary ComfortCuff NIPB, Pulse Oximetery, and predictive temperatures in one compact, easily transported monitor. Ideal for Spot Checks or continous vital signs monitoring.
Features
- DOX or Nellcor OxiMax SpO2 technology
- Proprietary ComfortCuff NIBP measures on inflation for faster, more comfortable readings
- Optional FILAC FasTemp predictive temperature with predictive and monitoring modes delivers quick, accurate measurements
- Audible and visual alarms provide instant notification
- Lightweight design and versatile mounting options—adapts readily to department and transport requirements
- Optional internal printer for demand, automatic interval, or trend reports.
Smart battery management with rechargeable sealed lead acid battery
Non-Invasive Blood Pressure
- Technique: Oscillometric measure upon inflation
- Measurement Time: <40 seconds typical; 0 to 300mmHg;
standard adult cuff
- Automatic Measurement
- Cycles: 1 - 60 minute; 2 - 4 hour; STAT
- Inflation Pressure Range: 30 to 300mmHg (Adult)
30 to 150mmHg (Pediatric)
20 to 150mmHg (Neonate)
- Resolution: 1mmHg
- Clinical Accuracy: SP-10:2002
- Transducer Accuracy: ±2mmHg or ±2% of reading, whichever is
greater, over full range
- STAT mode: 5 minutes of consecutive readings
Pulse Oximetry
- Range: 1% to 99%
- Resolution: 1%
- Accuracy (DOX only): 70% to 99% range: ±2%;
50% to 69% range: ±3%;
<50%: unspecified
- Accuracy (OxiMax Only): 70% to 99% range: ±2%;
<70%: unspecified
- Method: Dual wavelength LED
- Audio: Tone with each detected pulse; pitch varies
with saturation
- Heart (Pulse) Rate
- Range: 20 to 300bpm
- Accuracy (DOX only): Greater of ±1bpm or ±1%, of reading
Unspecified for 240 bpm
- Accuracy (OxiMax only): Greater of ±1bpm or ±1%, of reading
Unspecified for 240 bpm
- Update Time: Every second
- Resolution: 1bpm
- Pulse Volume: 10 levels or Off
Temperature (optional)
- Type: FILAC FasTemp
- Method: Predictive or Continuous
- Channels: 1
- Range: 30.0º to 43.0º C
- Accuracy: ±0.2º C for 30.0º to 36.9ºC;
±0.1º C for 37.0º to 39.0ºC;
±0.2º C for 39.1º to 43.0ºC
- Resolution: 0.1º F or 0.1º C
- Predictive Response Time: Oral: less than 10 seconds;
Axial: less than 10 seconds
- Rectal: less than 10 seconds
- Probes/Wells: Oral/Axial FILAC FasTemp [202026]
Rectal FILAC FasTemp (optional) [202034]
Isolation Well, Oral/Axial [202089]
Isolation Well, Rectal (optional) [202087]
Alarms
- Indication: Audible; Visual
- Volume: 10 levels or Off
- Silence: Yes; 2 minutes or permanent
Trend Memory
- Memory: Last 24 hours
- Parameters: SpO2, pulse rate, NIBP, temperature
- Report Types: Single Patient or Multi-patient; Selectable formats
Monitor Configurations:
506N3 - NIBP, Heart Rate
506NP3 - NIBP, Heart Rate, Printer
506NT3 - NIBP, Heart Rate, FILAC FasTemp™ Temperature
506NTP3 - NIBP, Heart Rate, FILAC FasTemp™ Temperature, Printer
506DN3 - DOX™ SpO2, NIBP, Heart Rate
506DNP3 - DOX™ SpO2, NIBP, Heart Rate, Printer
506DNT3 - DOX™ SpO2, NIBP, Heart Rate, Temperature
506DNTP3 - DOX™ SpO2, NIBP, Heart Rate, Temperature, Printer
506LN3 - OxiMax™ SpO2, NIBP, Heart Rate
506LNP3 - OxiMax™ SpO2, NIBP, Heart Rate, Printer
506LNT3 - OxiMax™ SpO2, NIBP, Heart Rate, Temperature
506LNTP3 - OxiMax™ SpO2, NIBP, Heart Rate, Temperature, Printer
506NV3 - NIBP, Heart Rate, TurboTemp™ Temperature
506NVP3 - NIBP, Heart Rate, TurboTemp™ Temperature, Printer
506DNV3 - DOX™ SpO2, NIBP, Heart Rate, TurboTemp™ Temperature
506DNVP3 - DOX™ SpO2, NIBP, Heart Rate, TurboTemp™ Temperature, Printer
506LNV3 - OxiMax™ SpO2, NIBP, Heart Rate, TurboTemp™ Temperature
506LNVP3 - OxiMax™ SpO2, NIBP, Heart Rate, TurboTemp™ Temperature, Printer
What's In The Box?
- Adult BP Cuff
- Large Adult BP Cuff
- NIBP Hose
- Power Cord (North American)
- Operators Manual
- Help Card
- Probe Covers (Box of 20) *Temp Units Only
- FasTemp Oral Temp Probe *Temp Units Only
- Isolation Well for Oral Probe *Temp Units Only
Download 506N3 Documentation PDF